Files
poona-ams/docs/ASSUMPTIONS.md
T

10 KiB

Assumptions Log — Poona Hospital AMS

Items marked "to be confirmed" or left flexible in the proposal are recorded here along with the configurable defaults implemented in Phase 1.

  1. Contract Expiry Alert Thresholds: Default alert lead time is 30 days (contract_expiry_alert_days), with critical (red) escalation at 7 days (contract_critical_alert_days). Both are admin-editable in Settings.
  2. Warranty & Insurance Expiry Lead Days: Default alert window is 30 days prior to end date (warranty_expiry_alert_days, insurance_expiry_alert_days), configurable by System Admin.
  3. PM & Calibration Due Windows: Planned PM and Calibration items transition from upcoming to due at 15 days before due_date / planned_date (calibration_due_window_days, pm_due_window_days), and become overdue once past the due date. Only completed (and rescheduled for PM) are persisted in the DB; upcoming/due/overdue are computed at read time against effective_today().
  4. Complaint SLA & Escalation Thresholds: Default target response/restoration SLA is 4 hours for critical priority (critical_complaint_sla_hours) and 24 hours for non_critical (non_critical_complaint_sla_hours), with unclosed critical complaints escalating after 24 hours (escalation_open_complaint_hours).
  5. Complaint Closure Remarks: require_closure_remarks defaults to true so engineers must document closure notes before closing a breakdown call; configurable in Settings.
  6. Asset Code Format: Asset codes follow <asset_code_prefix>-<DEPT_CODE>-<0001> with default prefix PHRC (asset_code_prefix), e.g. PHRC-ICU-0001.
  7. Default PM Service Mode: New PM schedules default to in_house (default_pm_mode, enum in_house | outsourced).
  8. Demo Equipment Return Alert: Expected return alerts trigger 3 days before expected_return_date (demo_return_alert_days), configurable in Settings.
  9. Demo Clock (demo_as_of_date): System-wide effective_today() reads demo_as_of_date (YYYY-MM-DD) from Settings if non-empty; otherwise uses the real current date in Asia/Kolkata.
  10. Soft Deletion for Masters & Users: Deleting departments, locations, categories, vendors, or users sets is_active = False rather than hard-deleting rows so historical asset/complaint/maintenance foreign keys remain intact.
  11. Password Hashing on Python 3.13: passlib==1.7.4 is incompatible with bcrypt>=4.1 on Python 3.13 (__about__ attribute removed / 72-byte wrap-check error), so app/core/security.py uses bcrypt (bcrypt==5.0.0) directly.

Assumptions added post-Phase-8

  1. Permission Specifics: system_admin and biomedical can create and edit full asset records (coordinator is restricted to updating requires_calibration, insurance_applicable, and remarks). Calibration frequency edits and plan approvals are restricted to system_admin, management, and coordinator (biomedical receives HTTP 403). department_user is scoped to their assigned department (ICU in demo data) across assets, complaints, PM, calibration, contracts, notifications, dashboard, reports, and global search, whereas system_admin, biomedical, management, and coordinator view all departments hospital-wide. management and coordinator cannot raise complaints (enforced server-side with HTTP 403).
  2. Critical Complaint Escalation & SLA Evaluation: APScheduler runs alerts.py daily (default 06:00 in configured TIMEZONE) and raises critical_complaint notifications when open critical complaint age exceeds escalation_open_complaint_hours (default 24h) or breaches compute_sla_status against critical_complaint_sla_hours (default 4h) / non_critical_complaint_sla_hours (default 24h).
  3. Insurance Drawer: Insurance summary is surfaced via a read-only slide-over panel (InsuranceDrawer) from AssetDetailPage, AssetPicker, and ComplaintDetailPage; creating and editing policies is handled via modal dialogs on /insurance and the Asset Insurance tab.
  4. Demo-Only Features: The TopBar Demo Clock (demo_as_of_date, admin-only date override for demonstrating expiry/overdue alerts and dynamic schedule transitions) and the deterministic seed loader (python -m app.seed.seed) exist specifically for demonstration and verification.
  5. Reopen Clears restored_at: Reopening a closed breakdown complaint clears both closed_at and restored_at (None) on the complaint row so downtime resumes as ongoing until re-restored.
  6. Immediate Asset Return to Service: When a breakdown complaint is restored or closed (and no other unrestored critical complaints exist on the asset), Asset.status transitions immediately to in_service rather than requiring an extra post-repair sign-off step.
  7. Code Prefixes (complaint_no_prefix & asset_code_prefix): Complaint numbers are auto-generated using complaint_no_prefix (default CMP, e.g. CMP-2026-0001; BD- or another prefix can be configured in Settings). Asset codes are entered manually to match physical hospital asset tags rather than auto-generated, so asset_code_prefix (default PHRC) acts as a configurable hospital label prefix shown in Settings and used for the bulk-import template sample code.
  8. Monthly Maintenance Uptime % Display: The monthly-maintenance backend report aggregates uptime/downtime in minutes per department rather than returning per-asset uptime_seconds/downtime_seconds, so Uptime % (uptime / (uptime + downtime) * 100, 1 decimal place) is computed on the frontend in ReportDetailPage.tsx (falling back to 100 - downtime_pct) and highlighted in the summary strip and table.
  9. Attachment & Data Retention Policy (§4.9, §8): In accordance with NABH digital hospital management standards and Indian statutory medical equipment documentation regulations, all equipment service history, breakdown complaint logs, preventive maintenance records, calibration certificates, and uploaded files (PDF/JPG/PNG up to 10 MB) are retained for a minimum mandatory period of 7 years from creation. Condemned and decommissioned equipment records are permanently maintained in read-only archive status and are never hard-deleted, ensuring unbroken historical auditability. Attachments are stored under backend/data/uploads/ with UUID filenames and served strictly through authenticated, role-gated download endpoints. Automated backups (backend/scripts/backup.py) bundle both the database and attachments into timestamped archives with configurable rolling retention (default: 7 copies).
  10. Master Enums vs Editable Tables (§8): Critical operational dimensions — complaint priority (critical, non_critical), complaint call_status (open, in_progress, closed), equipment status (in_service, under_repair, condemned, out_of_service), and contract types (AMC, CMC, OTHER) — are intentionally implemented as strict code-enforced Python/TypeScript enums rather than dynamically editable master tables. This design protects the business calculation engine (downtime/uptime percentages, SLA timers, asset status transitions) against data corruption or undefined states that arbitrary user renames could cause.
  11. Client Sign-Off & Default Thresholds (§4.3, §4.8, §5, §6): The hospital administration configures and approves operational thresholds via the /settings management console:
    • critical_complaint_sla_hours: default 4 hours.
    • non_critical_complaint_sla_hours: default 24 hours.
    • escalation_open_complaint_hours: default 24 hours.
    • contract_expiry_alert_days: default 30 days (red critical threshold: 7 days).
    • calibration_due_window_days: default 15 days.
    • pm_inhouse_lead_days: default 7 days; pm_outsourced_lead_days: default 15 days.
    • require_closure_remarks: default True.
    • downtime_milestone: default restoration (downtime terminates when equipment is restored).
  12. Alert Recurrence & Notification Re-trigger Policy (§6): The alert engine runs daily at 06:00 Asia/Kolkata and employs deterministic deduplication keys (dedupe_key) to prevent duplicate in-app notifications on consecutive days while a condition remains active. Overdue alerts escalate in severity to red, and an optional repeat reminder interval (alert_repeat_days, default 3 days) re-notifies coordinators and biomedical teams until the underlying task (PM, calibration, contract renewal, or breakdown) is addressed. SMTP email dispatch can be enabled in settings when hospital IT confirms mail server credentials.
  13. Duplicate Record & Import Conflict Prevention (Task 1):
    • Contracts: A duplicate contract is defined by (asset_id, contract_type, start_date, end_date) matching any existing contract in the database (including superseded ones). POST /api/contracts rejects exact duplicates with HTTP 409 (A contract with the same asset, type and dates already exists.). Contract renewals via POST /api/contracts/{id}/renew remain unaffected. In bulk import, in-file duplicate rows receive Duplicate of row N in this file. and rows duplicating database records receive Already exists in the system..
    • Insurance Policies: A duplicate insurance policy is defined by trimmed, case-insensitive policy_no matching any existing policy in the database. POST /api/insurance rejects duplicates with HTTP 409 (Policy number already exists.). In bulk import, in-file duplicate rows receive Duplicate of row N in this file. and rows duplicating database records receive Already exists in the system..
    • Calibration Plans: Each asset can have at most one calibration plan. In bulk import, rows for assets that already have a calibration plan in the database receive Asset already has a calibration plan., while rows repeating an earlier row for the same asset in the same upload file receive Duplicate of row N in this file.. Dry run previews surface these validation errors identically.
  14. Import Template Multi-Sheet Structure (Task 2): All four bulk-import templates (Assets, Contracts, Calibration Plans, Insurance Policies) feature an empty data template on Sheet 1 (containing only styled headers), an 'Example' reference on Sheet 2 (with warning banner 'Example only - this sheet is ignored on import' in Row 1 and sample values in Row 2), and 'Master Reference' lookup tables on Sheet 3. All four importers read strictly the first worksheet (wb.worksheets[0]), ensuring sample rows and reference sheets are never accidentally imported.